SPECIFICATION and FUNCTIONAL CAPABILITY by MODEL
Model Mini TOC
Sample volume flow 14ml/min
Detection limit 0,5 ppb
Accuracy +/- 0,5 ppb or 1%, whatever is greater
Range 0.5 - 1,000 ppb
Calibration calibration with standard and dilution water
Requirements for sample wate <2μS/cm ; <1bar and <50°C
Analysis time 2 sec
Pressure/temp max 1,5 bar (or pressure reducer), 50°C (or heat exchanger)
Powe 230 VAC (+ 10 %, - 15 %)
Dimensions(WxHxD) 250 x 500 x 200 mm
Weight ca. 14 kg
Control interface 8: TFT Touch Screen, 윈도즈 기반
UV Oxidation(partial) and Conductivity(difference) 1 Port for Sample, 1 for SST (Basis), 1 Port for Sample, 3 für SST (Professional)
Inputs
Interface USB, Analog output
Pressure reducer (toc free) to reduce pressure to 0,2 bar